CS-200 is an instrument that enables highly accurate luminance and chromaticity measurements, comparable to those of the spectroscopic type of instrument while maintaining the simplicity, operability, and competitive price of the tristimulus type.
The CS-200 features 40 sensors and performs calculations using the spectral sensitivity characteristics (colour-matching functions) corresponding to the sensitivity of human eye. With this newly developed spectral fitting method, tristimulus values (XYZ for red, green, and blue) with a far higher accuracy than those of conventional tristimulus colorimeters can be obtained, providing excellent accuracy.

Measurements over a wide range of luminance levels from a low luminance of 0.01 cd/m2 through high luminance of 20,000,000cd/m2 (with measuring angle of 0.1°) can be done.
Three measuring angles can be selected: 1°, 0.2°, and 0.1°. The measuring angle can be switched according to the measurement target, ranging from display devices or other objects with large areas to measure through objects with tiny areas to measure.
Technical Specifications
| Model |
CS-200 |
| Measurement range |
0.01 - 200,000 cd/m2 (Measuring angle 1°)
0.01 - 5,000,000 cd/m2 (Measuring angle 0.2°)
0.01 - 20,000,000 cd/m2 (Measuring angle 0.1°) |
Accuracy
(Measuring angle 1°) #1
(Temperature: 23°C±2°C, Relative humidity: 65% max.) |
150 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.002
0.01-0.5 cd/m2 ( for Illuminant A ) : Lv ±0.02cd/m2 ±1digit ; ---
0.5-1 cd/m2 ( for Illuminant A ) : Lv ±0.02 cd/m2 ±1digit ; xy ±0.007
1-10 cd/m2 ( for Illuminant A ) : Lv ±2 % ±1digit ; xy ±0.004
10-200,000 cd/m2 ( for Illuminant A) : Lv ±2 % ±1digit ; xy ±0.003
5000 cd/m2 ( for Illuminant A ) : color filter (R, G, B) ; xy ±0.006 |
Repeatability
(Measuring angle 1°) #2 |
0.01-1 cd/m2 ( for Illuminant A ) : Lv ±0.01 cd/m2 ±1digit ; --- (2σ/AUTO)
1-2 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.002 (2σ/AUTO)
2-4 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.001 (2σ/AUTO)
4-8 cd/m2 ( for Illuminant A ) : Lv ±0.5 % ±1digit ; xy 0.0005 (2σ/AUTO)
8-200,000 cd/m2 ( for Illuminant A ) : Lv ±0.1% ±1digit ; xy 0.0004 (2σ/AUTO) |
| Measurement time |
Automatic setting between 1s and 60s (AUTO); Automatic setting to 1s or 3s (LTD. AUTO)
0.5 sec / meas. (Super-FAST) 1 sec / meas.(FAST)
3 sec / meas. (SLOW) 12 sec / meas. (Super-SLOW) |
| Measurement method |
Spectral method, Grating + linear photo diode array |
| Measuring angle |
1°, 0.2°, 0.1°(switchable) |
| Minimum measuring area |
0.5 mm
0.1 mm (close up lens) |
| Minimum measuring distance |
296 mm (Distance from front edge of metal lens barrel) |
| Observer |
2/10 degrees |
| Color space |
Lv x y, Lv u' v', Lv TΔuv, XYZ, dominant wavelength |
| Measurement synchronization setting range |
Vertical synchronization frequency : 40.00 to 200.00Hz |
| Interface |
USB 1.1 |
| Power source |
AC adapter or 4 AA-Size Batteries |
| Battery life |
Approx. 3 hours (continuous measurement / Fast mode / AA-size alkaline cells) |
| Size |
95 mm (W) × 127 mm (H) × 330 mm (L) |
| Weight |
1.8 kg (without battery) |
| Operating temperature/humidity range |
0°C to 40°C, relative humidity 85% or less (at 35°C) with no condensation |
| Storage temperature /humidity range |
0°Cto 45°C,relative humidity 85% or less (at 35°C)with no condensation |
|